98.8 and 96.4 are 2.4 percentage points apart, which sounds small until you restate it as impurity. Janoshik is leaving 1.2 per cent of the detected area unassigned and Medu Laboratories 3.6 per cent — a factor of 3 between them. The impurity fraction is the quantity that moves when a method changes, and it is the one worth arguing about; the headline is just its complement. Comparable means same column chemistry, same gradient slope, same detection wavelength, same integration convention. Until you have those four from both laboratories, 98.8 and 96.4 are two measurements of slightly different quantities that happen to share a unit.
Read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.
Mass shifts and what they usually mean
| Δ mass (Da) | Most likely cause | Distinguishing feature |
|---|
| +1 | Deamidation (Asn or Gln) | New peak, slightly earlier retention |
| −17 | Loss of ammonia | Often with deamidation |
| −18 | Dehydration / succinimide | pH-dependent, reversible |
| +16 | Oxidation (Met, Trp) | Earlier retention, light-related |
| −128 | Missing Gln or Lys | Deletion sequence from synthesis |
| 0 | Isomer: racemisation or scrambling | Same mass, shifted retention |
The relevant detail is that buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
If you are ranking vendors, specify a method and have all samples tested at the same place.